Product | Easidew PRO I.S. for Gases | Easidew PRO LQ I.S. for Liquids |
Performance Specifications |
Measurement Range | -110 to +20°C (-166 to +68°F) dew point; -100 to +20°C (-148 to +68°F) dew point | 0 to 1000 ppmW capability – factory configured to customer-required range and application |
Accuracy | ±1°C (±1.8°F) dew point (+20 to -60°C / +68 to -76°F) ±2°C (±3.6°F) dew point (-60 to -110°C / -76 to -166°F) |
Response Time | 5 mins to T95 (dry to wet) |
Repeatability | 0.5°C (0.9°F) dew point |
Calibration | Traceable 13-point calibration and certificate |
Electrical Specifications |
Output Signal | 4-20 mA (2-wire connection, current source); User configurable over range |
Output | Dew point or moisture content | Moisture content |
Analog Output Scaled Range | Dew point:-110 to +20°C (-166 to +68°F) or –100 to +20°C (-148 to +68°F)Moisture content in gas:0-3000 ppmV Non-standard:mg/m3, lbs/MMSCF natural gas | Moisture content in liquid: 0 to 1000 ppmW capability – factory configured to customer-required range and application |
Supply Voltage | 12 to 28 V DC |
Load Resistance | Max 250 Ω @ 12 V (500 Ω @ 24 V) |
Current Consumption | 23 mA max, depending on output signal |
Saturation Constants (for moisture in liquids measurements only) | | 6-point look-up table for saturation constants up to 1000 ppmW over the temperature range 0 to +50°C (32 to +122°F); Saturation constants for 10 common liquids can be programmed into the Easidew PRO XP IS via the application software; Alternatively the user can program saturation constants manually |
CE Conformity | Approved |
Operating Specifications |
Operating Temperature | –40 to +60°C (-40 to +140°F) |
Compensated Temperature Range | -20 to +50°C (-4 to +122°F) NOTE: The transmitter accuracy statement is only valid for the temperature range: -20/+50°C (-4/+122°F) |
Storage Temperature | -40 to +60°C (-40 to +140°F) |
Operating Pressure | 45 MPa (450 barg / 6526.7 psig) maximum |
Flow rate
| 1 to 5 Nl/min mounted in standard sampling block 0 to 10 m/sec direct insertion | 0.1 to 0.3L/min through Easidew sample block 0.1 to 1m/s direct insertion |
Mechanical Specifications |
Ingress Protection | IP66 in accordance with standard BS EN 60529:1992; NEMA 4 protection in accordance with standard NEMA 250-2003 |
Intrinsically safe area certificates* | ATEX: II 1 G Ex ia IIC T4 Ga (–20°C to +70°C)
IECEx: Exd ia IIC T4 Ga (–20°C to +70°C)
TC TR 0Ex ia IIC T4 Ga (–20°C to +70°C)
CCSAUS: Class I, Division I Groups A B C D, T4
FM: Class I, Division 1, Groups A B C D, T4 |
Russian Pattern Approval | Russia (GOST-R), Kazakhstan (GOST-K) |
Canadian Pressure Vessel Certification | C.R.N. — 11 Canadian provinces |
Oxygen service
| Optional: cleaned for enriched oxygen | |
Housing material | 316 stainless steel |
Filter (sensor protection) | Standard: Stainless steel sintered guard (for protection against fine particulate >80μm) Optional: HDPE guard (for protection against fine particulate >10μm) |
Process Connection and
Material | 316 stainless steel 5/8” — 18 UNF with bonded seal, G1/2” and 3/4” UNF adaptors available
(material certificate to BS EN 10204 3.1 — see ‘accessories and spare parts’ on product order codes) |
Weight | 1.27kg (2.8lb) |
Electrical Connections | Screw terminal via female M20 x 1.5mm gland |
Diagnostic conditions (factory programmed) | Condition Output
Sensor fault 23 mA Under-range dew point 4 mA Over-range dew point 20 mA |
Approved galvanic isolators | KFD2-CR-EX1.20200 KFD2-CR-EX1.30200 KFD0-CS-EX2.50P KFD0-CS-EX1.50P KFD2-STC4-EX1.H MTL5040 MTL5541 MTL5041 |