Performance | |
Measurement Technology | Thermo-paramagnetic oxygen sensor |
Measured Gas | Oxygen (O2) |
Background Gas | Analyzer is calibrated in the background gas of the process. Standard backgrounds include N2, CO2, CH4, H2, He & BIOGAS. |
Gas Requirements | Non-condensing sample with particles <3μm |
Measurement Range | Analyzer is calibrated to the range the customer requires. Select from: 0...0.5 % up to 0...50 % and 20, 80 or 90...100 % |
Display Resolution | 0.01 % 0.1 % for suppressed zero ranges |
Display Type | Backlit LCD (not on 501-GP2 model) |
Intrinsic Error (Accuracy) | < ±1 % of range or ±0.02 % O2, whichever is greater < ±0.2 % O2 or 1 % of range, whichever is greater (for suppressed zero ranges 20/80/90...100 %) |
Response Time (T90) with High Speed Response (HSR) Enabled | < 15 seconds |
Repeatability | ±0.2 % of range or 0.02 % O2, whichever is greater |
Linearity | ±0.5 % of range or 0.05 % O2, whichever is greater |
Zero Stability | ±0.25 % of range per month |
Range Stability | ±0.25 % of range per month |
Sample Flow Rate | 100...500 ml/min (0.2...1.27 scfh) |
Sample Flow Effect (Calibrated at 300 ml/min) | < 1 % of range for flows within stated range |
Sample Pressure | 0.75...2 Bar A (unit must be calibrated at the same pressure as sample) |
Sample Temperature | A constant temperature between 5 and 55 °C |
Sample Cell Temperature | +50 °C |
Electrical Specifications | |
Analog Inputs | 1 off mA input for an external sensor (can be displayed on the screen) 1 off mA input to act as an active compensation for the process conditions |
Analog Outputs | 2 off 4...20 mA outputs (powered with 24V excitation voltage) |
Output Ranges | Primary range is set to the calibrated range of the instrument. The second is user-selectable |
Alarms | 2 off single pole changeover relays for concentration (250 V, 5 A max) |
Datalogging | Digital communications can be used to log the output from the analyzer on an external device |
Digital Communications | Modbus RTU over RS485 |
Power Supply | 24 V DC, 1.5 A max |
Operating Conditions | |
Ambient Temperature | +5...+40 °C |
Mechanical Specifications | |
Warm-Up Time | < 25 minutes (at 20 °C ambient) |
Stabilization Time | 5 minutes |
Wetted Materials | 316 stainless steel, borosilicate glass, platinum, (plus O-ring) |
Dimensions (w x d x h) | 260 x 180 x 128 mm |
Weight | Approx. 3 kg |
O-Ring Materials | Viton |
Gas Connection | 1/8” bulkhead standard |
Ingress Protection | IP55 |