Performance | |
Measurement Technology | Zirconia (ZR) |
Measurement Range | 0 to 1,000ppm / 0 to 25% / 0 to 96% |
Output Resolution ( For %) | 0.01% |
Output Resolution (For Ppm) | 1ppm |
Accuracy | +/-2% of reading (or 2ppm O2) @ calibrated temperature and pressure |
Response Time (T90) | <10 seconds @ 20˚C. Warm up time: 15mins from switch on. |
Ldl (Sensitivity) | 0.01% (when measuring %) / 1ppm (when measuring ppm) |
Temperature Range | -20˚C to +50˚C |
Pressure Range | Regulated to 5 psi(G) |
Linearity | 2% of reading |
Life Expectation | >30,000 hours in 20.9% oxygen (ambient air) |
Humidity | 0-95% RH non-condensing |
Shelf Life | No shelf life |
Electrical Input / Output | |
Power Supply | Internal rechargeable battery and charger (Optional) Charge via 3-pin IEC Mains connector @ 85-240 VAC 50-60H |
Power Consumption | Max Load 160mA |
Signal Output | 4-20mA |
Digital Communications | RS232 bi-directional Modbus protocol. |
Connection Ports | Sample In: 6mm Sample Exhaust Out: Via Flowmeter port. |
Display Format | % O2 / ppm O2 |
User Interface | LCD wiith 4 button multifunction keypad |
Mechanical Specifications | |
Dimensions | See technical drawings |
Weight | <4.5kg / 4.5kg / 8.5kg |
Case Materials | High impact copolymer polypropylene |
Sample Gas | Standard 0.5 SLPM |
Internal Regulation | Sample pressure & flow set at 5PSIG / 0.3BARG; 0.55SCFH / 250 ml/m |
Ingress Protection | IP67 when case closed for transportation |
Sample Connection | 6mm |
Certification | Complies with EMC Directive 2004 / 108 / EC. UL/ETL Certification Number: UL-61010-1 |