Measurement Technology | Choice of PED, Zirconia, Electrochemical, Metal Oxide Moisture Sensor, Quartz Crystal |
Impurities Detected | Combination or choice of N2, O2, H2O, CnHm |
Sample Gas | PED: Ar- HE Other sensors: multiple gases |
| For full sensor specifications download the LD8000 MultiGas Datasheet |
Operating Temperture Range | +5 to +45 °C |
Sample Gas Temperture | 0 to +100 °C |
Sample Flow Requirement | Sensor dependent: see datasheet |
Operating Sample Pressure Range | 3-30 psig (for lower sample pressure requirement, an additional high purity pump is used) |
Outlet Pressure | Atmospheric |
Inlet Fittings | 1/8'' or 1/4'' Swagelok compression or VCR |
Outlet Fittings | 1/8'' or 1/4'' Swagelok compression or VCR |
Standard Features | Manual or autoranging, MCU embeded system, touchscreen 5.6'' LCD, 4-20mA outputs, alarm historic, digital outputs for status, 3 ranges, calibration in progress |
Options | Internal sampling system for zero/span/sample, serial communication RS232-RS485-Modbus-Profibus, 2 level alarms, zero gas free calibration system |
Supply | 110VAC 50-60Hz / 220VAC 50-60Hz |
Power Consumption | 100-250 watts depending of the combination of sensors and options mounted in the unit |
Enclosure Type | 3U or 4U rackmount type depending of the combination of sensors and options mounted in the unit |
Ingress Protection | IP20 in accordance with IEC 60529 |
Enclosure Finish | RAL7030 powder coat |
Weight | 25-40 lbs (11-18 kg) depending on the combination of sensors and options mounted in the case |
Certification | In compliance with EMC directive 2004/108/EC, EN 61000-6-2:2005 for immunity & EN 61000-6-4:2007 for emissions |