Hydrocarbon Dew-Point Measurement |
Measuring Technique | Dark SpotTM fixed sample analysis. Direct photo-detection of hydrocarbon condensate at hydrocarbon dew-point temperature |
Sensor Cooling |
Automatic via 3-stage Peltier effect
electronic cooler under adaptive control |
Maximum Range |
Up to Δ -55K measurement depression from Main Unit operating temperature |
Accuracy |
±0.5 °C hydrocarbon dew point (single and multiple condensible component analysis) |
Sample Flow |
0.03 m³/hr (0.5 Nl/min) - alarm standard |
Measurement Frequency |
6 cycles/hour (recommended) 12 cycles/hour (maximum) |
Operating Pressure |
15…100 barg (218…1450 psig) |
Water Dew-Point Measurement (Optional) |
Measuring Technique |
Michell Ceramic Metal-Oxide Moisture Sensor |
Units - Moisture content |
°C and °F water dew point; lb/MMscf; mg/m3, ppmV |
Resolution |
0.1 °C, 0.1 °F |
Range |
Calibrated from -100 to +20 °C dew point |
Accuracy |
±1 °C from -59 to +20 °C dew point
±2 °C from -100 to -60 °C dew point |
Sample Flow |
1 Nl/min (0.06 m3/hr) – alarm standard |
Temperature Coefficient |
Algorithm compensation (-20...+50 °C) |
Operating Pressure |
30…138 barg (435…2000 psig) |
Pressure Measurement Hydrocarbon Dew-Point Analysis Pressure
Optional Water Dew-Point Process Pressure
|
Units |
MPa, barg, psig |
Resolution |
0.1 MPa and barg, 1 psig |
Range |
HC dew point: 0...100 barg W dew point: 0...200 barg |
Accuracy |
±0.25 % FS |
Hydrocarbon Dew-Point Analyzer |
Resolution |
Hydrocarbon and water dew point: 0.1 °C |
Sample Gas Supply |
Natural gas up to 100 barg, pressure regulated in sampling system |
Enclosure |
EExd cast enclosure with removable glass window viewing port. Internally heated for condensation protection |
Sample Gas Connections |
1/8” NPT(F) for both hydrocarbon and water dew-point channels; Sampling system: 6 mm OD or 1/4” OD |
Operating Environment |
Indoors/Outdoors -20...+50 °C; Max 95 %rh |
Power Supply |
90...260 V AC, 50/60 Hz, 125 W Main Unit 300 W c/w Sampling System (indoor) 400 W c/w Sampling system (outdoor) |
Weight |
22.5 kg Main Unit 60 kg (approx) c/w Sampling System (indoor) 75 kg (approx) c/w Sampling system (outdoor) |
Integrated Display/Keyboard |
Touch screen with vacuum fluorescent display |
Outputs |
Modbus RTU, RS485 @ 9600 baud rate Two 4...20 mA linear (non-isolated) outputs, user-configurable for any
combination of dew-point or pressure parameters |
Alarms |
Process and analyzer status via software register and display annotation Integrated low flow alarms for each sample
flow Analyzer status fault flag 23 mA on mA Output 1 |
Certification |
Hazardous Area Certification |
ATEX/UKCA: II 2G Ex db IIB + H2 T3 Gb Tamb -40°C...+60°C)
IECEx: Ex db IIB + H2 T3 GbTamb -40°C...+60°C)
cQPSus: Class I, Division 1, Groups B, C and D
T3 Tamb -25°C...+60°C)
CL I ZONE 1: Ex db IIB+H2 T3 Gb, AEx db
IIB+H2 T3 Gb
Tamb -40°C...+60°C |
Additional |
TRCU 012 (EAC), PESO (India), NEPSI (China), Ukraine Ex |
Pattern Approval |
GOST-R, GOST-K |